In the Resistant Starch Assay Kit (K-RSTAR), will weighing less Pancreatic alpha amylase (< than 1g/100mL) affect the results?

Modified on Tue, 9 Nov, 2021 at 12:59 PM


We recommend performing the assay exactly as described in the method. 


Lower PAA weights may be weighed but the volume of buffer must also be adjusted equally, for example 1g/100mL or 0.5g/50mL.


Changing this ratio may affect the analysis and any changes to this method will need to be validated internally.